Recall Z-1595-2023 ยท MEC Disinfector, Medical Devices

Steris Corporation

Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062

DateMay 17, 2023 (initiated Apr 10, 2023)
StatusOpen, Classified
ReasonIn the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE FIELD CORRECTION dated 3/28/23 was sent to customers. STERIS Action: All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their unit(s) to perform the correction. User Action: Users can continue to use their Reliance Vision Multi-Chamber Washer/Disinfector prior to completion of the correction. Should your unit alarm Drying Temperature Too High, abort the cycle, turn off the power to the unit, and contact STERIS. The competent (regulatory) authority of your country has been informed of this notice. Customer res
Quantity322 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of: China, Lithuania, Saudi Arabia, Korea, Republic of, Canada, Taiwan, United Kingdom, Spain, Australia, New Zealand, Spain, Brazil, Bahrain, Switzerland, Great Britain, South Africa, Hong Kong, Qatar.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.