Recall Z-1462-2023 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

MAQUET CARDIOSAVE hybrid Intra-Aortic Balloon Pump, Model Number 0998-XX-0800-32

DateApr 26, 2023 (initiated Apr 4, 2023)
StatusOpen, Classified
ReasonThe power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle.
Root causeComponent design/selection
ActionAn URGENT FIELD SAFETY NOTICE MEDICAL DEVICE CORRECTION was sent to customers beginning 4/4/23. Actions to be taken by the Customer now: Our records indicate that that you have received one or more CARDIOSAVE hybrid IABPs having the serial numbers that are affected by this recall. 1. Inspect the IABP to determine if it has a Type J power plug (refer to Image 1 below). If present, the IABP should not be used until the cord reel has been replaced. 2. Schedule a service appointment with your Datascope Corp./Getinge Service Representative for the cord reel correction from a Type J connect
Quantity54 units
DistributionInternational distribution in the country of Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112372CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP
Cardiac Assist, Maquet Cardiovascular, LLC, Sep 15, 2011