Recall Z-1470-2023 ยท IZL System, X-Ray, Mobile

FUJIFILM Healthcare Americas Corporation

FujiFilm FDR AQRO (Model DR-XD1000)- A digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams

DateApr 27, 2023 (initiated Feb 27, 2023)
StatusOpen, Classified
ReasonArm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.
Root causeDevice Design
ActionFUJIFILM Healthcare Americas Corporation (Fujifilm) issued Urgent Medical Device Recall letter on 2/27/23 via email to Distributors ad end-Users. Letter states reason for recall, health risk and action to take: A Fujifilm service engineer will contact your facility to arrange a visit to repair your affected unit(s). ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1)Perform an inspection according to the steps outlined below prior to using the product. a. If there are no abnormalities present during the pre-use inspection the product can continue to be used safely. b. Immediately remove the device
Quantity77 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170858FDR AQRO (DR-XD 1000)
Fujifilm Medical Systems U.S.A, Inc., Apr 24, 2017