ClearanceBrief
Recall Z-1656-2023 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Maquet Medical Systems USA

QUADROX-iD Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMOD 70000-USA 701067840 BEQ-HMOD70000-USA 701067859

DateMay 30, 2023 (initiated May 18, 2023)
StatusOpen, Classified
ReasonProduct packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
Root causeUnder Investigation by firm
ActionOn May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 opti
Quantity24,250 (US)
DistributionNationwide domestic distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101153QUADROX-ID ADULT DIFFUSION MEMBRANE OXYGENATOR WITH SOFTLINE COATING, MODEL: BE-HMOD 70000-USA, BEQ-HMOD 70000-USA, HMOD
Maquet Cardiopulmonary, AG, Dec 8, 2010