ClearanceBrief
Recall Z-1464-2023 ยท GCJ Laparoscope, General & Plastic Surgery

Covidien LP

VersaOne Reusable Positioning Cannula 11mm Standard, Model Number RC11STS

DateApr 26, 2023 (initiated Mar 22, 2023)
StatusOpen, Classified
ReasonThere is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated March 2023 was sent to customers. Customer Actions: Our records show that your facility has received the impacted product. Medtronic requests that you take the following actions: " Identify and quarantine any used and unused impacted product. Please refer to Attachment A: Impacted Serial Number Lists for a list of affected products. For verifying the respective product and serial number on the device and packaging please refer to Attachment B. " Sterilize all used devices as per your standard decontamination protocol prior to returning
Quantity51 units
DistributionWorldwide - US Nationwide distribution in the states of NC and the countries of Belgium, Denmark, France, Germany, India, Italy, Netherlands, Spain, Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201805VersaOne Reusable Positioning Trocar System
Covidien, Aug 26, 2020