Recall Z-1600-2023 ยท QJQ Cranial Electrotherapy Stimulator To Treat Insomnia And/Or Anxiety

Fisher Wallace Laboratories Inc.

Fisher Wallace Stimulator Models FW-100 and FW-200

DateMay 17, 2023 (initiated Apr 6, 2023)
StatusOpen, Classified
ReasonIf the batteries are installed incorrectly, the product may overheat, possibly leading to pain, discomfort, irritation, or even burns during use of the device.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE CORRECTION dated 4/6/23 was sent to customers. Short term actions to be taken by the User: 1. Always ensure the batteries are inserted correctly into the device (see above diagram). 2. Do not leave batteries in the device when the device is not in use (e.g. in between a morning and evening treatment). 3. Do not fall asleep while using the device because corrective actions cannot be taken when the user is asleep. 4. If the device exhibits excess heat, burning smell, or odor, turn off the device, and discontinue use. Contact Fisher Wallace Laboratories immediately. The
Quantity43,489 total devices
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K903654LISS CRANIAL STIMULATOR MODEL SBL202-B
Medical Consultants Intl. , Ltd., Oct 4, 1990