Recall Z-1477-2023 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America Llc

Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and fluoroscopy (R/F) system

DateApr 28, 2023 (initiated Oct 21, 2022)
StatusOpen, Classified
ReasonPotential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking
Root causeComponent design/selection
ActionPhilips issued a revised customer letter on 3/10/23 via U.S. mail -URGENT Medical Device Correction Notification Update CombiDiagnost R90 and ProxiDiagnost N90 Component damage due to short circuit & table stop due to broken tabletop cable. Customers received previous customer communications : 21 Oct 2021 for ProxiDiagnost N90 1.0 (706100); 17 Nov 2022 for ProxiDiagnost N90 1.1 (706110). Circulate this notice to all users of the device, attach the notice as an addendum to the IFU (User Manual). If a user observes fire in the M-cabinet, sees or smells smoke, or hears a popping noise comin
QuantityTotal Quantity: 234 units. US 195units. R.O.W 39 units
DistributionWorldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Chile, China, Germany, India, Italy, Malaysia, Netherlands, Panama, Spain, Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173433ProxiDiagnost N90
Philips Medical Systems Dmc GmbH, Feb 5, 2018
K212837ProxiDiagnost N90
Philips Medical Systems Dmc GmbH, Sep 21, 2021