Recall Z-1426-2023 · OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

SPINEART SA

JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S

DateApr 24, 2023 (initiated Mar 16, 2023)
StatusOpen, Classified
ReasonJULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.
Root causeProcess change control
ActionOn 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of
Quantity544
DistributionUS Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190877Juliet® Ti LL Lumbar Interbody Device
Spineart, May 29, 2019