Recall Z-1414-2023 ยท IZL System, X-Ray, Mobile

NeuroLogica Corporation

Samsung GM85 Digital X-ray Imaging System-A Digital Diagnostic Mobile X-ray System

DateApr 19, 2023 (initiated Mar 8, 2023)
StatusOpen, Classified
ReasonIssue related to the operation of the arm latch within the column that supports the arm, potential bodily harm due to a fall of the arm
Root causeDevice Design
ActionNeuroLogica (Importer)for Samsung Electronics GM85 digital x-ray system issued Urgent Medical Device Correction letter issued 3/8/23 via email. Letter states reason for recall, health risk and action to take: Please identify the system label and confirm whether or not your device is one of those affected by this FSN. The list of affected serial numbers is provided in the Appendix. Samsung Electronics will undertake the voluntary corrective action considering the possibility of potential bodily harm due to the fall of the arm. Service engineers will correct all affected devices free of charge
Quantity935 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K222353GM85
Samsung Electronics Co., Ltd., Sep 29, 2022