Baxter Healthcare Corporation

GEM NEUROTUBE, Absorbable Woven Polyglycolic Acid Mesh Tube, Product Codes: a) 5311-01240-010, b) 5311-01240-012; for peripheral nerve repair

DateApr 25, 2023 (initiated Mar 27, 2023)
StatusOpen, Classified
ReasonThe product is brittle and potentially crumbling upon handling or when being removed from its package.
Root causeUnder Investigation by firm
ActionBaxter issued an Urgent Medical Device Recall notice to its consignees on 03/27/2023 by mail (USPS). The notice explained the issue and the risk and requested the following actions: 1. Immediately locate, isolate, and cease all use of the affected lot number of the product. 2. Contact Synovis MCA to arrange for return and replacement of the product. Synovis MCA can be reached at 800-510-3318 or 205-941-0111 between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday. Please have your Synovis MCA account number, product code, lot number, and quantity of product to be retur
Quantity169 units
DistributionWorldwide distribution - US Nationwide and the countries of China, Germany, Sweden, Norway, Netherlands, Greece, Switzerland, France