ClearanceBrief
Recall Z-1471-2023 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Draeger Medical Systems, Inc.

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

DateApr 27, 2023 (initiated Apr 5, 2023)
StatusOpen, Classified
ReasonThe software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Root causeSoftware design
ActionAn Urgent - Medical Device Correction notification letter dated March 2023 was sent to customers. Actions to be taken: Until the corrected software is available and installed, clinicians should utilize the Draeger ICS/MSOO Instructions For Use while applying any additional means of standard practices for assessment of patients with unusually high and/or low QRS amplitudes. Clinicians should combine this information with physical assessment of the patient and follow standard clinical practice. Please ensure that all users of the affected products listed above and other people within your orga
Quantity6803 systems
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151860Infinity CentralStation Wide
Draeger Medical Systems, Inc., Sep 2, 2015