Recall Z-1429-2023 ยท CBK Ventilator, Continuous, Facility Use

Hamilton Medical AG

HAMILTON-C6, REF: 160021

DateApr 25, 2023 (initiated Mar 14, 2023)
StatusOpen, Classified
ReasonSoftware error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.
Root causeSoftware design
ActionOn 3/14/23, correction notices were emailed to health care facilities and distribution partners. Customers were told the following: 1) Firm will reach out to schedule the software update. 2) See Operation's Manual, Section 7.7, Safety Ventilation. The device must turn off to exit Safety Ventilation. 3) Complete and return the confirmation form to bret.everett@hamiltonmedical.com 4) If the malfunction occurs take the following actions: a) Monitor your patient. b) Provide an alternative device for ventilation. c) Have the ventilator serviced. If you have questions or concerns, contact Techni
Quantity103
DistributionUS Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201658Hamilton-C6
Hamilton Medical AG, Dec 11, 2020