Recall Z-1381-2023 ยท OAB Low Energy Direct Current Thermal Ablation System

Angiodynamics, Inc.

NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN: H787204001110

DateApr 12, 2023 (initiated Mar 2, 2023)
StatusOpen, Classified
ReasonNot programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure
Root causeUnder Investigation by firm
ActionAngioDynamics issued Urgent Medical Device Recall Letter on March 2, 2023 via Federal Express. Letter states reason for recall health risk and action to take: IMMEDIATELY Stop using the product subject to recall. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location. Segregate this product in a secure location for return to AngioDynamics, Inc. Forward a copy of this recall notification to all sites to which you have distributed affected product. 2. ComplFx Reply Verification Tracking Form: Attn: NK Probe 5-Pa
Quantity10 boxes( 5 probes/box)
DistributionUS Nationwide Distribution: CA,CO, FL, IA, NY, OH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183385NanoKnife System
Angiodynamics, Jun 18, 2019