Recall Z-1572-2023 ยท MID System, Test, Anticardiolipin Immunological

Bio-Rad Laboratories, Inc.

BioPlex 2200 REF 665-2050, APLS IgM Pack, APLS IgM Reagent Pack

DateMay 10, 2023 (initiated Feb 28, 2023)
StatusOpen, Classified
ReasonAPLS IgM reagent kits were packaged with the incorrect conjugate, which could lead to an increase in false-positive and false-negative results.
Root causeUnder Investigation by firm
ActionOn February 28, 2023, Bio-Rad Laboratories issued an "Urgent: Medical Device Recall" letter to affected consignees. In addition, to informing consignees about the recall, they ask consignees to take the following actions: We are requesting that customers perform quality control using the BioPlex APLS IgM Control Set (Catalog No. 663-2030, UDI GTIN 00847865000642) with each reagent pack ( single pack validation ) of BioPlex 2200 APLS IgM Reagent Pack, Lot No. 301538, until replacement product is provided. If QC fails (controls below or above the acceptable range), please discontinue use of t
Quantity932 Reagent Packs
DistributionUS Nationwide distribution including in the states of AL, AR, AZ, CA, CT, FL, HI, IL, IN, KS, KY, LA, MI, MO, NC, ND, NE, NJ, NM, NY, OH, OK, PR,SC, TN, TX, UT, VT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130528BIOPLEX 2200 APLS IGM
Bio-Rad Laboratories, Oct 21, 2013