ClearanceBrief
Recall Z-1390-2023 ยท GAL Suture, Absorbable, Natural

Covidien, LP

Plain Gut Absorbable Sutures, PLAIN GUT 5/0 18 P-13 X12 (SG-5686), PLAIN GUT 4-0 18 UD SC-11 DA (SG828-2)

DateApr 12, 2023 (initiated Mar 1, 2023)
StatusOpen, Classified
ReasonMedtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.
Root causeProcess control
ActionAn "URGENT PRODUCT RECALL" letter dated 3/1/23 was sent to customers. Actions to be Taken: " Immediately identify and quarantine all unused affected product (Refer to Attachment A for identifying affected product). " Please complete and return the enclosed Customer Confirmation Form, even if you have no product to return and email to rs.gmbfcamitg@medtronic.com. " Product purchased directly from Medtronic please contact rs.covidienfeedbackcustomerservice@medtronic.com for Return Goods Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return
Quantity3228 units
DistributionWorldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142656Surgical Gut Suture
Covidien, Jan 29, 2015