Recall Z-1281-2023 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs

DateMar 22, 2023 (initiated Feb 17, 2023)
StatusOpen, Classified
ReasonIt is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
Root causeUnder Investigation by firm
ActionA Field Safety Notice, Medical Device Correction letter dated 2/16/23 was sent to customers. Actions to be taken by the user " Be aware that an RT plan exported by using the VSIM Export button will always have the DICOM attribute Beam Limiting Device Angle (300A,0120) set to zero. " Be aware that an RT plan exported by using the VSIM Export button will always have the DICOM attribute Number of Wedges (300A,00D0) set to zero and no wedge information will be exported. " Be aware that an RT plan exported by using the VSIM Export button will never include the DICOM attribute Applicator Sequ
Quantity10770 systems
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190387RayStation
RaySearch Laboratories AB (PUBL), Jun 19, 2019