Recall Z-1356-2023 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America Llc

ProxiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 706100 and 706110

DateApr 10, 2023 (initiated Oct 21, 2021)
StatusOpen, Classified
ReasonIf the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.
Root causeDevice Design
ActionAn IMPORTANT PRODUCT NOTICE dated 10/21/21 was sent to customers with the CombiDiagnost R90 system. The system may experience a short circuit of the Electromagnetic Interference (EMI) filter or UA/UB board, with a potential for smoke and/or a popping sound coming from the Main cabinet. If a user sees or smells smoke or hears a popping noise coming from the M-cabinet, immediately remove power to the room (main circuit breaker) and remove the patient. Stop using the system for further examinations. An IMPORTANT PRODUCT NOTICE dated 10/21/21 was sent to customers with the ProxiDiagnost N90.
Quantity243 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov