ClearanceBrief
Recall Z-1355-2023 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America Llc

CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031

DateApr 10, 2023 (initiated Oct 21, 2021)
StatusOpen, Classified
ReasonIf the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.
Root causeDevice Design
ActionAn IMPORTANT PRODUCT NOTICE dated 10/21/21 was sent to customers with the CombiDiagnost R90 system. The system may experience a short circuit of the Electromagnetic Interference (EMI) filter or UA/UB board, with a potential for smoke and/or a popping sound coming from the Main cabinet. If a user sees or smells smoke or hears a popping noise coming from the M-cabinet, immediately remove power to the room (main circuit breaker) and remove the patient. Stop using the system for further examinations. An IMPORTANT PRODUCT NOTICE dated 10/21/21 was sent to customers with the ProxiDiagnost N90.
Quantity512 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163210Philips CombiDiagnost R90
Philips Medical Systems Dmc GmbH, Jan 31, 2017
K203087CombiDiagnost R90
Philips Medical Systems Dmc GmbH, Dec 3, 2020