Recall Z-1310-2023 ยท FRN Pump, Infusion

Iradimed Corporation

Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,

DateMar 30, 2023 (initiated Feb 24, 2023)
StatusOpen, Classified
ReasonCertain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.
Root causeProcess control
ActionIradimed Corporation issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/24/2023 by certified mail. The notice explained the issue and the risk and provided updated instructions for use to mitigate the potential risk of inlet occlusion. For further information concerning this issue or have any questions, please contact Iradimed support via email at SyringeAdapterRecall@Iradimed.com or via telephone at 407-677-8022.
Quantity3285 boxes of 50 devices
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South