Recall Z-1315-2023 ยท KCY Tourniquet, Pneumatic

Stryker Sustainability Solutions

Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-235; Color Cuff Single Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-135; Color Cuff Dual Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-235; Color Cuff Single Por

DateApr 1, 2023 (initiated Mar 1, 2023)
StatusOpen, Classified
ReasonIncreased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.
Root causeProcess control
ActionOn 3/1/23, recall notices started being distributed to customers. Customers were asked to do the following: 1) Follow the IFU and inspect for tears. Do not use the device if any damage is noted. Return the device and packaging to the recalling firm, if it is not in acceptable condition for surgery. 2) To avoid a pinched or unseated O-ring upon connection of the tourniquet cuff to the tourniquet controller, hold down the connector release on the tourniquet controller and take extra caution to ensure connection between the cuff and controller is not at an angle. 3) If a cuff does not reach or
Quantity127,894
DistributionUS: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC, VT, CT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.