Recall Z-1332-2023 ยท MBP Filler, Bone Void, Osteoinduction (W/O Human Growth Factor)

Medtronic Sofamor Danek USA Inc

Medtronic Accelerate Grafton DBF a) ACCELERATE BG SET T50203 GRAFTON DBF 3CC, REF T50203; b) ACCELERATE BG SET T50206 GRAFTON DBF 6CC, REF T50206; c) ACCELERATE BG SET T50209 GRAFTON DBF 9CC, REF T50209; d) ACCELERATE BG SET T50212 GRAFTON DBF12CC, REF T50212; Demineralized Bone Matrix (DBM) product

DateApr 4, 2023 (initiated Feb 8, 2023)
StatusOpen, Classified
ReasonPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Root causeProcess control
ActionMedtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.
Quantity3037 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechte
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.