Recall Z-1418-2023 ยท MND Ligator, Esophageal

Olympus Corporation of the Americas

Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model HX 201UR 135 - Rotatable, 230 cm long, 2.8 scope channel, standard clip Model HX 201UR 135L - Rotatable, 230 cm long, 2.8 mm scope channel, Long clip

DateApr 20, 2023 (initiated Feb 23, 2023)
StatusOpen, Classified
ReasonThere have been complaints that the clip did not come out of the tube sheath during the procedure.
Root causeProcess change control
ActionThe firm, Olympus, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated 2/23/2023 to its customers. The letter described the product, the issue and actions to be taken. The customers were instructed to do the following: 1) immediately assess, cease use of and quarantine any affected (lot numbers) products you have identified. 2.) Call your Olympus customer service representative at 1-800-848-9024, option 2 to receive a RMA to return any affected product at not charge. 3.)Access the Olympus recall portal to indicate that you have received this notification at https://olympusamerica.com/recall.
Quantity65.9 boxes (373 pieces)
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013066HX-5/6-1 ENDOSCOPIC CLIPPING DEVICE
Olympus America, Inc., Dec 11, 2001