Recall Z-1328-2023 ยท MBP Filler, Bone Void, Osteoinduction (W/O Human Growth Factor)
Medtronic Sofamor Danek USA Inc
Medtronic Grafton Matrix Strips: a) DBM S42275 8MMX1CMX10CM 2EA MATRIX STRPS, REF S42275; b) DBM S42280 8MMX1CMX20CM 2EA MATRX STRPS, REF S42280; c) DBM T42275 8MMX1CMX10CM 2 EACH GRAFTON M, REF T42275; d) DBM T42275AUS 8MMX1CMX10CM 2 EA GRFTON S, REF T42275AUS; e) DBM T42275INT 8MMX1CMX10CM 2 EA GRFTON S, REF T42275INT; f) DBM T42275JPN GRAFTON MATRIX 1CM X 10CM, REF T42275JPN; g) DBM T42
| Date | Apr 4, 2023 (initiated Feb 8, 2023) |
| Status | Open, Classified |
| Reason | Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier. |
| Root cause | Process control |
| Action | Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed. |
| Quantity | 23850 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechte |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K051195 | GRAFTON DBM Osteotech, Inc., Dec 16, 2005 |