Recall Z-1478-2023 ยท FBK Endoscopic Injection Needle, Gastroenterology-Urology

Labories Medical Technologies

injeTAK Adjustable Tip Needles Needle Sheath diameter 6Fr with 35-50cm in length Stainless-stell needle cannula is 25 gauge

DateApr 28, 2023 (initiated Feb 7, 2023)
StatusOpen, Classified
ReasonThe packaging may be damaged, compromising the sterile barrier.
Root causeProcess control
ActionA customer letter dated February 7, 2023 was issued to impacted consignees. The letter asks consignees to examine their inventory for affected product lots and to complete a response card to the recalling firm. Return of impacted product will be coordinated with Laborie Medical Technologies.
Quantity415,480 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090830LABORIE INJETAK ADJUSTABLE TIP NEEDLE, FLEXIBLE TIP AND RIDGID TIP
Laborie Medical Technologies, Ltd., Jul 21, 2009