Recall Z-1320-2023 ยท DYB Introducer, Catheter

Biosense Webster, Inc.

Biosense Webster Carto Vizigo Bi-Directional Guiding Sheath, 8.5F(Medium), REF D-1385-02-S, RX CE 2797, Sterile

DateApr 1, 2023 (initiated Feb 21, 2023)
StatusOpen, Classified
ReasonIncorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch
Root causeUnder Investigation by firm
ActionOn March 2, 2023, Biosense Webster, Inc. issued an "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" notification to affected US customers via FedEx. The notification was issued to foreign consignees on February 29, 2023. In addition to informing consignees about the recall Biosense Webster asked customer to take the following actions: 1. Carefully review the information contained in this Medical Device Correction/Removal. 2. Ensure that anyone in your facility who needs to be aware of this notification reads the attached letter carefully. 3. Evaluate if you have inventory of CARTO VIZIGO" Bi-Direc
Quantity316 units
DistributionWorldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WI. The country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170997CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath
Biosense Webster, Inc., Jul 25, 2017