Recall Z-1235-2023 ยท JAD System, Therapeutic, X-Ray

Sensus Healthcare, Inc.

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

DateMar 8, 2023 (initiated Feb 10, 2023)
StatusTerminated
ReasonWhen the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.
Root causeSoftware design
ActionSensus healthcare issued an "Urgent Device Correction" to its consignees on 02/10/2023 by email. The notice requested the following actions: "Do not initiate beam delivery after a pause & resume sequence without verifying the amount of dose delivered. Do evaluate any patient treatments with version 1.9.2 that followed the pause & resume workflow. Do verify that the amount of dose to be delivered is as expected after each pause & resume sequence." The firm has developed a permanent software resolution for this issue. Customer service will contact the customer to coordinate upgrading to this
Quantity23 units
DistributionUS Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150037Sensus Healthcare Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities