ClearanceBrief
Recall Z-1348-2023 ยท MQL Media, Reproductive

CooperSurgical, Inc.

Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060

DateApr 6, 2023 (initiated Feb 13, 2023)
StatusOpen, Classified
ReasonIt has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.
Root causeLabeling Change Control
ActionAn URGENT: VOLUNTARY MEDIA RECALL dated 1/26/23 was sent to customers. Actions to be Taken: For both CUSTOMERS and DISTRIBUTORS: " Inspect your stock for the Product from this lot. Lot information can be found on the box and bottle labeling. o If found, quarantine, and discontinue use of the Product from this lot immediately. o Please complete and return the appropriate enclosed Acknowledgment Form (even if you do not have the specific lot remaining in stock) to initiate our recall return process. Credit will be applied to your account to reorder replacement product, as per our recall pro
Quantity190 units
DistributionWorldwide - US Nationwide distribution in the states of NY, UT and the countries of Argentina, Belgium, Bulgaria, Ecuador, France, Germany, Hong Kong, Italy, Malaysia, Mexico, Peru, Singapore, Spain, Switzerland, Thailand, Turkey, the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142991LifeGlobal Global Total w/HSA, LifeGlobal Global Total for Fertilization w/HSA, LifeGlobal Global Total w/HEPES W/HSA
Life Global Group, LLC, Feb 17, 2015