Recall Z-1282-2023 ยท LPL Lenses, Soft Contact, Daily Wear

CooperVision, Inc.

Clariti 1-day toric soft contact lenses, Power -1.00 -1.25/90

DateMar 23, 2023 (initiated Feb 14, 2023)
StatusOpen, Classified
ReasonThe lenses manufactured in the affected lot were made with a misaligned axis resulting in lenses with the incorrect power.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE CORRECTION letter dated 2/14/23 was sent to customers. Eye Care Practitioner/Distributor Instructions: Our records indicate you have received product from the affected lot. We request you take the following actions: " Please examine your inventory/product shelf for any of the provided affected lot number. The lot number can be found on the back panel of the carton below the expiration date. Below is an example and where to locate the lot number. " Stop any further distribution of product from the designated lot number; " We recommend you contact your Eye Care Practiti
Quantity3150
DistributionWorldwide - US Nationwide distribution in the states of FL, IL, NY, UT, and WI. The country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181920Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker
CooperVision, Inc., Dec 11, 2018