ClearanceBrief
Recall Z-1411-2023 · KWP Appliance, Fixation, Spinal Interlaminal

Medtronic Sofamor Danek USA Inc

INFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

DateApr 19, 2023 (initiated Feb 16, 2023)
StatusOpen, Classified
ReasonIncorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Root causeUnder Investigation by firm
ActionThe firm issued an URGENT MEDICAL DEVICE RECALL NOTICE to its consignees on 02/16/2023 by UPS 2nd day mail. The letter explained the issue and the risk and requested the following actions be taken: Immediately identify, segregate, and quarantine affected products within your inventory. Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Contact Medtronic to return affected product: 1-800-933-2635 to receive an return authorization (RGA) and coordinate inventory replacement/credit.
Quantity2 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163375INFINITY™ OCT System
Medtronic Sofamor Danek USA, Inc., Aug 21, 2017