Recall Z-1338-2023 ยท OMA Amplitude-Integrated Electroencephalograph

Moberg Research, Inc.

Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements Model: CNS-350

DateApr 4, 2023 (initiated Feb 20, 2023)
StatusOpen, Classified
ReasonBattery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC
Root causeUnder Investigation by firm
ActionNatus issued Letters Urgent Medical Device Recall Letter Component Neuromonitoring System (Moberg CNS Monitor Model CNS-350) via first class beginning Feb 20, 2023. Letter states reason for recall, health risk and action to take: Ensure that the CNS is either shut down or plugged into a power outlet. Perform the next step while using proper personal protective equipment (PPE), including gloves. On the top of the CNS-350 device, open each of the three (3) battery ports and remove any backup battery. 2) Inspect the battery and the battery port for any white, brown, or green/blue buildup, or for
Quantity89 units
DistributionWorldwide distribution - US Nationwide and the countries of France, Germany, Italy, Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080217COMPONENT NEUROMONITORING SYSTEM
Moberg Research, Inc., Nov 5, 2008