Recall Z-1346-2023 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE

DateApr 5, 2023 (initiated Mar 1, 2023)
StatusOpen, Classified
ReasonSome Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 3/1/23 was sent to customers. RECOMMENDED ACTIONS Terumo CVS advises users to remove all affected product from inventory and return unused product to Terumo CVS following the instructions provided in this notice. Terumo CVS will credit the user s accounts for returned, unused products. AFFECTED POPULATION The affected population is all patients on which an impacted Capiox NX, RX or FX Oxygenator is used. CUSTOMER INSTRUCTIONS - Review this Medical Device Recall notice and ensure that all users receive notice of this issue. Refer t
Quantity84 units
DistributionUS Nationwide. Belgium, Canada, Costa Rica, Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180950Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation), Terumo Capiox NX19 Oxygenator with Reservoir (west Orie
Terumo Cardiovascular Systems Corporation, Jun 22, 2018