Recall Z-1275-2023 ยท QLG Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems

Abbott Diabetes Care, Inc.

FreeStyle Libre Reader, REF: 71951-01, 71952-01, 71953-01 a component of the FreeStyle Libre 2 Flash Glucose Monitoring System

DateApr 6, 2023 (initiated Feb 13, 2023)
StatusOpen, Classified
ReasonLithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.
Root causeDevice Design
ActionOn 02/13/2023, correction notices were mailed or emailed to customers, health care professionals, and distribution partners who were asked to do the following: To determine if the Reader needs to be replaced, answer these questions: Does Reader appear damaged, cracked, or swollen? Has Reader been too hot to hold? Is it no longer able to hold a charge? If NO to all, perform a Reader Test. See www.FreeStyleBattery.com If yes or if the test failed, discontinue use and call Customer Service for replacement. - Don't expose the Reader, power adapter or yellow USB cable to liquids. - Store the Read
Quantity1,534,532
DistributionU.S. Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193371FreeStyle Libre 2 Flash Glucose Monitoring System
Abbott Diabetes Care, Inc., Jun 12, 2020