Recall Z-1286-2023 ยท DCK C-Reactive Protein, Antigen, Antiserum, And Control

Randox Laboratories

CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

DateSep 30, 2024 (initiated Feb 1, 2023)
StatusOpen, Classified
ReasonThe Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
Root causeUnder Investigation by firm
ActionRandox UK (manufacturer) extended Urgent Medical Device Removal letter to the US distributor via email on 2/1/23. The distributor contacted the customer via email on 2/6/23. Letter states reason for recall, health risk and action to take: " Discontinue use of and discard any of the above immediately. Provide Randox with photographic evidence of the destruction of the kits. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the pa
Quantity2 kits
DistributionPA, OH, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953212C-REACTIVE PROTEIN IMMUNOTURBIDIMETRIC KIT & CRP CALIBRATOR
Randox Laboratories, Ltd., Jul 31, 1995