ClearanceBrief
Recall Z-1421-2023 ยท HDF Speculum, Vaginal, Metal

Carefusion 2200 Inc

(1) Carefusion V. Mueller Graves Vaginal Speculum, Catalog #GL11, Medium Blade, 4 x 1 3/8" (10.2 X 3.5cm); and (2) Carefusion V. Mueller Pederson Vaginal Specula, Catalog #GL31, Medium, 4 x 7/8" (10.2 x 2.2cm).

DateApr 21, 2023 (initiated Mar 16, 2023)
StatusOpen, Classified
ReasonA portion of the two products were incorrectly marked (etched) and packaged resulting in a product mismatch. The labels and laser etchings were switched on the two products.
Root causeLabeling mix-ups
ActionThe recalling firm issued two different letters dated 3/16/2023 via FedEx 2-day and email, one to distributors and one to other consignees that were for the attention of Medical Director, Risk Manager, Medical Device Safety Officer, Lab Manager, Clinical Engineering, etc. (medical facilities). The letter to the medical facilities explained the issue, provided drawings of the two different catalog numbers and included a photograph of one of the products, provided the clinical impact, and provided the actions to be taken. These actions were to check their inventory locations for the affected p
Quantity726 units
DistributionDistribution was made to AL, AR, AZ, CA, CO, CT, FL, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, VA, WA, WI, and WV, including Puerto Rico. There was also government/military distribution. Foreign distribution was made to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.