Recall Z-1323-2023 ยท MBP Filler, Bone Void, Osteoinduction (W/O Human Growth Factor)

Medtronic Sofamor Danek USA Inc

Medtronic Grafton Putty: a) DBM S43101 GRAFTON 0.5CC PUTTY, REF S43101; b) DBM S43102 GRAFTON 1CC PUTTY, REF S43102; c) DBM S43103 GRAFTON 2.5CC PUTTY, REF S43103; d) DBM S43105 GRAFTON 5CC PUTTY, REF S43105; e) DBM S43110 GRAFTON 10CC PUTTY, REF S43110; f) DBM T43101 0.5CC GRAFTON PUTTY, REF T43101; g) DBM T43101INT 0.5CC GRAFTON PUTTY, REF T43101INT; h) DBM T43102 1CC GRAFTON PUTTY

DateApr 4, 2023 (initiated Feb 8, 2023)
StatusOpen, Classified
ReasonPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Root causeProcess control
ActionMedtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.
Quantity186889 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechte
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051195GRAFTON DBM
Osteotech, Inc., Dec 16, 2005