Recall Z-1191-2023 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Stryker GmbH

stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved

DateMar 1, 2023 (initiated Feb 7, 2023)
StatusOpen, Classified
ReasonStryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.
Root causeProcess change control
ActionThe firm issue an URGENT MEDICAL DEVICE RECALLL its consignees by letter on 02/07/2023. The notice explained the product and the issue and requested the following actions be taken: 1. Check inventory for affected product and remove from use 2, Sign and return the enclosed Business Reply Form by email to fieldaction@stryker.com to confirm receipt of this notification/documenting product disposition. Stryker will will contact you to arrange for the return of the affected device(s) 3. Spread awareness of this communication internally until all required actions have been completed within your
Quantity488 units
DistributionUS Nationwide distribution.