ClearanceBrief
Recall Z-1268-2023 · MRR System, Test, Low Density, Lipoprotein

Siemens Healthcare Diagnostics, Inc.

Atellica CH LDL Cholesterol Direct (DLDL)-In vitro diagnostic quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin) using the Atellica CH Analyzer. SMN: 11097632

DateAug 29, 2023 (initiated Jan 27, 2023)
StatusOpen, Classified
ReasonReagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
Root causeUnder Investigation by firm
ActionSiemens Issued Urgent Field Notice to US customers ACHC23-01.B.US via FedEx beginning on 01/30/2023. In accordance with outside US country regulations, Siemens in-country personnel for the EU and non-EU regions follow local requirements for contacting their respective National Competent Authorities customers for distribution. and regional reporting to their local competent authorities beginning on 01/27/2023. The UMDC/UFSN provides instructions to all Atellica CH 930 customers to either separate the assays causing carryover by analyzer, if they have multiple analyzers, or by batch testing o
Quantity8420 units
DistributionNationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Guadeloupe Hong Kong Hungary India Ireland Israel Italy Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982529RANDOX DIRECT LDL CHOLESTEROL
Randox Laboratories, Ltd., Sep 11, 1998