Recall Z-1194-2023 ยท JAF Monitor, Ultrasonic, Nonfetal

Unetixs Vascular, Inc.

MultiLab Series II LHS TI & ERGO non-invasive vascular diagnostic device, Model Number 11949-0000-01

DateMar 1, 2023 (initiated Feb 10, 2023)
StatusOpen, Classified
ReasonThe current luer fittings and adaptors allow for the possibility of the air line to be misconnected to an intravenous line, which may cause an air embolism.
Root causeUse error
ActionUrgent Medical Device Notification letters were sent to customers beginning 2/10/23. In order to physically prevent device misconnections from occurring, standardized connector designs for specific medical applications have been developed in accordance with ISO-80369-5. Unetixs Vascular has adopted this standard and ceased shipment of devices, hoses and blood pressure accessories equipped with luer fittings. All new devices will ship with bayonet connectors. These new connectors are designed to help reduce the risk for medical device misconnection. Action Items: 1. Immediately order one Bayo
Quantity2310 units
DistributionWorldwide distribution - US Nationwide and Global distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K904392MULTILAB 2000
Unetixs, Inc., Apr 29, 1991