Recall Z-1260-2023 ยท KNW Instrument, Biopsy

Bard Peripheral Vascular Inc

Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK

DateMar 16, 2023 (initiated Feb 28, 2023)
StatusOpen, Classified
ReasonThe external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.
Root causeMixed-up of materials/components
ActionOn 2/28/23, "URGENT MEDICAL DEVICE RECALL" notices were mailed and emailed to customers and distributors who were told to take the following actions: 1) Discontinue use of and Dispose of all affected product remaining in your possession in accordance with your local facility's process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3) If you purchased this product from a distributor, contact your distributor for further instructions and credit resolution. 4) Complete and return the attached Customer Res
Quantity38,635
DistributionWorldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO, PA, TX, NC, CA, ME, MO, NV, MI, IL, NY, VA, AZ, NH, LA, WI, KY, MA, WA, CT, SD, OK, DC, IA, AR, GA, NE, NM, IN, NJ, MN, PR, AL, AK, SC, KS, OR, ID, VT, MT, WV, ND, TN, MS, HI, WY, RI and OUS countries of: Canada, Korea, Thai
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171953Bard Mission Disposable Core Biopsy Instrument
Bard Peripheral Vascular, Inc., Sep 14, 2017