ClearanceBrief
Recall Z-1227-2023 ยท GGP Test, Qualitative And Quantitative Factor Deficiency

Aniara Diagnostica LLC

BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K dependent, which inhibits coagulation. Protein C in plasma is measured following a specific activation with Protac (Agkistrodom Contortrix), an enzyme extracted from snake venom. Activated protein C hydrolysis the chromogenic substrate (SaPC-21)

DateMar 6, 2023 (initiated Feb 1, 2023)
StatusOpen, Classified
ReasonNew warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility
Root causeOther
ActionAniara issued Medical Field Correction Letter (FSCA#29) prepared by HYPHEN BioMed and emailed to the customer by Aniara Diagnostica on 2/1/23. Letter states reason for recall, health risk and action to take: New labelling will be implemented on all new lots as follow: x A new pictogram added on reagent and box label: GHS08 x Signal Word: Warning x Hazard Statements: H361f - Suspected of damaging fertility x Precautionary Statements: o P201 : Obtain special instructions before use. o P280 : Wear protective gloves/protective clothing/eye protection/face protection. o P308 + P313 : IF exposed or
Quantity2 units
DistributionUS Nationwide distribution in the state of NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050365BIOPHEN PROTEIN C5, C2.5, MODELS 221205, 221202
Hyphen Biomed, Dec 27, 2005