Recall Z-1201-2023 ยท OUG Medical Device Data System

Biomerieux Inc

MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments for ID/AST and blood culture activities and the laboratory information system (LIS).

DateMar 2, 2023 (initiated Feb 6, 2023)
StatusOpen, Classified
ReasonFor users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.
Root causeSoftware design
ActionBioMerieux, US subsidiary issued an URGENT FIELD SAFETY NOTICE to its consignees beginning on 02/06/2023 by email and letter. The notice explained the issue with the product and potential risk, and requested the use take the following actions: "1. Confirm this letter has been distributed to and reviewed by all appropriate personnel within your organization. 2. If you are using VITEK MS as part of your laboratory workflow AND you currently use MYLA AST Filter rules disable current MYLA AST Filter rules. Review AST results to determine if some results should be suppressed from reporting. 3. I
Quantity19 systems
DistributionUS Nationwide distribution to the states of Alabama, California, North Carolina, and Puerto Rico.