Recall Z-1272-2023 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Vapotherm

Low Flow Disposable Patient Circuit for the Precision Flow System, Product Code PF-DPC-LOW

DateMar 17, 2023 (initiated Dec 16, 2022)
StatusOpen, Classified
ReasonThere have been reports of excessive condensation associated with certain lots of the Low Flow Disposable Patient Circuits (DPCs) used with Precision Flow System which may lead to excessive rainout.
Root causeNonconforming Material/Component
ActionA Technical Bulletin dated 10/27/22 was distributed to customers. Vapotherm has learned of certain Low Flow Disposable Patient Circuit (DPC) lots exhibiting excessive condensation. Please find the affected DPC lots in the attached appendix. You have been identified as a customer that received some or all of these lots. This condensation can be mitigated by ensuring that the delivery tube is fully primed and warm before placing the therapy on the patient, per existing instructions for use. Always check that the delivery tube is warm across the entire length before placing the therapy on the
Quantity5869 total
DistributionWorldwide distribution - US Nationwide and the country of UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111640PRECISION FLOW(R) - HELIOX
Vapotherm, Inc., Oct 7, 2011