Recall Z-1185-2023 ยท PBZ Image Processing Device For Estimation Of External Blood Loss

Stryker Corporation

Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 3L Catalog Number: FG 12022. Intended adjunct in the estimation of blood loss.

DateFeb 27, 2023 (initiated Jan 25, 2023)
StatusOpen, Classified
ReasonTriton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.
Root causeUnder Investigation by firm
ActionStryker issued Urgent Medical Device Correction letter to Risk Manager, OR Director, Materials Manager dated Jan 25, 2023. Letter states reason for recall, health risk and action to take: Steps: 1. Immediately check your internal inventory to locate any affected Finished Goods Inserts and Scanning labels (FG 12010 and FG 12022). Quarantine and discontinue use of all affected products. 2. A Stryker representative will contact your facility to assist updating the Triton Canister software. 3. Return the enclosed Business Reply Form (BRF) even if you don t have any affected product on hand, via em
Quantity541 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142801Triton Canister System
Gauss Surgical, Inc.,, Mar 12, 2015