ClearanceBrief
Recall Z-1188-2023 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Maquet Medical Systems USA

BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0, Product Code 70106.9078. Part of the CARDIOHELP System, a pre-connected set used for extracorporeal respiratory and/or cardiovascular support.

DateFeb 28, 2023 (initiated Feb 13, 2023)
StatusOpen, Classified
ReasonFirm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.
Root causeProcess control
ActionOn February 13, 2023, the firm notified customers via Urgent Medical Device Correction letters. Customers were informed of the potential packaging issues and the associated risks to health. Customers were instructed to stop use and return all unused and unexpired affected product for credit. If customers have no alternative product, they may choose to continue use of the product, but they must perform a visual inspection of the primary packaging, checking for visible stress marks or damage to the packaging. If there are visible stress marks in the packaging, the customer should not use the p
Quantity10,587 units (US); 94,987 (worldwide)
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czechia, Denmark, Ecuador, Eg
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112360QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIES
Maquet Cardiopulmonary, AG, Sep 15, 2011