Recall Z-1210-2023 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

DiaSorin Molecular LLC

Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.

DateMar 3, 2023 (initiated Jan 30, 2023)
StatusOpen, Classified
ReasonDirect amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Root causeDevice Design
ActionOn 01/30/2023, correction notices were distributed to customers via in-person site visits. Customers were asked to do the following: 1) Discontinue use of affected devices using affected assay definitions. 2) Customers can contact the recalling firm to replace affected kits. The firm can also provide assistance with inventory reconciliation/destruction, and if needed, assistance with potential instrument/lab decontamination. 3) Any sample result with an error code (EC510, EC515, RNA IC, Insufficient Specimen Volume) should be retested following the Instructions for Use. 4) If leakage is suspe
Quantity4142
DistributionUS Nationwide distribution including in the states of NJ, ME, ND, IL, MI, FL, GA, AZ, MA, CA, MO, SD, MN, PA, OH, VT, NE, WI, CO, DC, VA, TX, NH, IN, IA, MD, NC, AL, NY, KY, WV, RI, KS, NM, OK, WA, SC, CT, TN, NV, LA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201505Simplexa Flu A/B & RSV Direct Gen II, Simplexa Flu A/B & RSV Positive Control Pack
Diasorin Molecular, LLC, Aug 22, 2020