Recall Z-1376-2023 ยท CAN Regulator, Pressure, Gas Cylinder

Western/Scott Fetzer Company

Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital VIPR System (with E-size cylinder) MNDR-600, Western Oxytote DTE Digital VIPR (head only) MNDS-602, Western Oxytote DTE Digital VIPR System (with D-size cylinder) MNDS-603, Western Oxytote DTE Digital VIPR System (with E-size cylinder)

DateApr 12, 2023 (initiated Jan 3, 2023)
StatusOpen, Classified
ReasonGauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
Root causeUnder Investigation by firm
ActionOn January 3, 2023, the firm issued a Safety Alert to customers. Customers were provided with updated operating instructions. Customers instructed to take the following immediate actions: 1. Implement updated filling and venting instructions which include steps for the filler to verify proper gauge pressure after vent down and after fill/re-fill. 2. Implement an additional verification of gauge pressure upon delivery at the healthcare facility. 3. Take care to minimize any impact to the cylinder during use, transportation, or processing during fill/re-fill. The firm continues to evaluate
Quantity24,287 units
DistributionUS Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
FDA recordOpen on fda.gov