ClearanceBrief
Recall Z-1158-2023 ยท LRS I.V. Start Kit

Cardinal Health 200, LLC

Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A

DateFeb 22, 2023 (initiated Nov 18, 2022)
StatusOpen, Classified
ReasonPresource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
Root causeNonconforming Material/Component
ActionCardinal Health issued Urgent: Product Correction letter and Acknowledgement Form was sent overnight on November 18, 2022 to the direct consignees. Letter states reason for recall, health risk and action to take: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming you
Quantity36 EA
DistributionUS Nationwide distribution in the states of GA, PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.