Recall Z-1209-2023 ยท QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
DiaSorin Molecular LLC
Simplexa COVID-19 Direct, REF: MOL4150; including Reaction Mix, REF: MOL4150; with previous assay definition v2
| Date | Mar 3, 2023 (initiated Jan 30, 2023) |
| Status | Open, Classified |
| Reason | Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes. |
| Root cause | Device Design |
| Action | On 01/30/2023, correction notices were distributed to customers via in-person site visits. Customers were asked to do the following: 1) Discontinue use of affected devices using affected assay definitions. 2) Customers can contact the recalling firm to replace affected kits. The firm can also provide assistance with inventory reconciliation/destruction, and if needed, assistance with potential instrument/lab decontamination. 3) Any sample result with an error code (EC510, EC515, RNA IC, Insufficient Specimen Volume) should be retested following the Instructions for Use. 4) If leakage is suspe |
| Quantity | 90142 |
| Distribution | US Nationwide distribution including in the states of NJ, ME, ND, IL, MI, FL, GA, AZ, MA, CA, MO, SD, MN, PA, OH, VT, NE, WI, CO, DC, VA, TX, NH, IN, IA, MD, NC, AL, NY, KY, WV, RI, KS, NM, OK, WA, SC, CT, TN, NV, LA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.