ClearanceBrief
Recall Z-1138-2023 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)

DateFeb 24, 2023 (initiated Jan 23, 2023)
StatusOpen, Classified
ReasonThe magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
Root causeError in labeling
ActionAn Urgent Medical Device Correction notification letter dated January 2023 was sent to customers. Actions to be taken: Use of the mask is contraindicated for patients and other persons in close vicinity to patients using the masks, that have implanted devices that may be affected by magnets, including but not limited to: Pacemakers Implantable cardioverter defibrillators (ICD) Neurostimulators Magnetic metallic implants/electrodes/valves placed in upper limbs, torso, or higher (i.e., neck and head) Cerebral spinal fluid (CSF) shunts (e.g., ventriculoperitoneal (VP) shunt) Aneurysm clips
Quantity893 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081743CLASSICSTAR NV SE AND NOVASTAR NV SE AND AAV FULL FACE MASKS
Draeger Medical AG & Co. KG, Sep 22, 2008