ClearanceBrief
Recall Z-1154-2023 · JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

Siemens Healthcare Diagnostics, Inc.

ADVIA Centaur Folate 500 test kit (REF)-For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10331250 Siemens Material Number (SMN): 10325366

DateFeb 21, 2023 (initiated Jan 9, 2023)
StatusOpen, Classified
ReasonNegative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
Root causeUnder Investigation by firm
ActionSiemens Healthineers issued Urgent Medical Device Corrections (UMDC AIMC 23-03.A-1.US and AIMC 23-03.A-2.US) to US customers and Urgent Field Safety Notices (UFSN AIMC 23-03.A-1.OUS and AIMC 23-03.A-2.OUS) to customers outside the United States. The UMDCs and UFSNs advise customers to use the appropriate FolSerum calibration instructions when testing serum samples with Atellica IM Folate kit lots ending in 337 and above or ADVIA Centaur Folate kit lots ending in 336 and above. Letter states reason for recall, health risk and action to take: Follow the instructions in this Urgent Medical Devic
Quantity124 kits
DistributionNationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cura¿ao,¿St¿Eus Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010050ADVIA CENTAUR AND ACS: 180 FOLATE IMMUNOASSAY
Bayer Diagnostics Corp., Feb 5, 2001